Description
HIOKI BT3562A Battery HiTester
Fully Automated Production Line Testing of Large Cells for xEVs or Mid-sized Packs of up to 100V
Key Features
- Simultaneous measurement of internal resistance and open circuit voltage.
- Fully automated production line testing of large cells for xEVs or mid-sized packs of up to 100V.
- Measure from 0Ω to 3.1kΩ (minimum resolution 0.1μΩ)
- Resistance measurement ranges: 3mΩ/30mΩ/300mΩ/3Ω/30Ω/300Ω/3kΩ
- Measure from 0V DC to ±100V DC (minimum resolution 10μV)
- Voltage measurement ranges: 6V/60V/100V
- Equipped with LAN
Features
Measuring battery performance and safety using internal resistance (AC-IR) and open-circuit voltage (OCV)
Inspect the quality of completed cells, modules, and packs on production lines. Measure internal resistance (AC-IR) and open-circuit voltage (OCV) to check battery quality. Cells produced at the cell production factory are shipped to the module production factory after undergoing a shipping inspection. Since factors such as vibrations during shipment and even the passage of time can cause defects, batteries undergo an acceptance inspection before being assembled into modules and packs.
BT3561A
Small cells for power motors, Small packs of up to 60V
Voltage measurement ranges: 6V/60V
Resistance measurement ranges: 30mΩ/300mΩ/3Ω/30Ω/300Ω/3kΩ
BT3562A
Large cells for xEVs, Mid-sized packs of up to 100V
Voltage measurement ranges: 6V/60V/100V
Resistance measurement ranges: 3mΩ/30mΩ/300mΩ/3Ω/30Ω/300Ω/3kΩ
BT3563A
Large packs for xEVs, Large packs of up to 300 V
Voltage measurement ranges: 6V/60V/300V
Resistance measurement ranges: 3mΩ/30mΩ/300mΩ/3Ω/30Ω/300Ω/3kΩ
Reducing test system development cost and management man-hours
The new BT356xA has dramatically improved tolerances for line resistance compared to previous models. This improvement makes it easy to build test systems with large numbers of channels using relays. Additionally, a longer maintenance cycle for systems in use means fewer maintenance man-hours. Finally, its capability to handle thinner cables than with previous models* makes it easier to route cables.
*AWG 29 (0.064 mm2) wire equivalent to 2.2Ω over an 8 m round trip can be used with the 3mΩ or 30mΩ range.
Improve Productivity by Increasing the Number of Inspection Channels
You can use the Switch Mainframe SW1001/SW1002 to increase the number of measurement channels. Additionally, you can perform scan measurement by controlling two instruments at once, for example a BT356xA series instrument and a DM7276, or a BT4560 and a DM7276.
Preventing instrument malfunctions caused by static electricity
Batteries can become charged on production lines, for example, when being transported on a conveyor belt. When probes are placed in contact with such batteries, the resulting application of static electricity can then damage the instrument. The BT356xA series is designed to withstand contact with ±30kV of static electricity*, preventing static-caused malfunctions and reducing testing line downtime.
* ±30kV IEC 61000-4-2 contact discharge
LAN interface as standard
The BT356xA series is equipped with a LAN interface as standard equipment, making it easy for the instrument to interoperate with a PLC*-based control system. The ability to use readily accessible LAN cables helps lower costs during system development and maintenance. Furthermore, a design with strong noise and static electricity resistance helps avoid system problems.
*Programmable Logic Controller, a device that automatically controls one or more machines
Specifications
Resistance Measurement Ranges | 3mΩ(Max. display: 3.1000mΩ, resolution: 0.1μΩ, measurement current: 100mA) 30mΩ(Max. display: 31.000mΩ, resolution: 1μΩ, measurement current: 100mA) 300mΩ(Max. display: 310.00mΩ, resolution: 10μΩ, measurement current: 10mA) 3Ω(Max. display: 3.1000Ω, resolution: 100μΩ, measurement current: 1mA) 30Ω(Max. display: 31.000Ω, resolution: 1mΩ, measurement current: 100μA) 300Ω(Max. display: 310.00Ω, resolution: 10mΩ, measurement current: 10μA) 3kΩ(Max. display: 3.1000kΩ, resolution: 100mΩ, measurement current: 10μA) Basic accuracy: ±0.5% rdg ±10 dgt(3mΩ range), ±0.5% rdg ±5 dgt(30mΩ range or more) |
---|---|
Voltage Measurement Ranges | 6V(Max. display: 6.00000V, resolution: 10μV) 60V(Max. display: 60.0000V, resolution: 100μV) 100V(Max. display: 100.000V, resolution: 1mV) Basic accuracy: ±0.01% rdg. ±3 dgt. |
Response Time | 10 ms |
Sampling Period | Ω or V (60 Hz): 4ms (EX.FAST), 12ms (FAST), 35ms (MEDIUM), 150ms (SLOW) ΩV (60 Hz): 8ms (EX.FAST), 24ms (FAST), 70ms (MEDIUM), 253ms (SLOW) Ω or V (50 Hz): 4ms (EX.FAST), 12ms (FAST), 42ms (MEDIUM), 157ms (SLOW) |
Functions | Contact check, Zero adjustment (±1000 counts), Pulse measurement, Comparator (Hi/ IN/ Lo), Statistical calculations (Max. 30,000), Delay, Average, Panel saving/loading, Memory storage, LabVIEW® driver |
Interfaces | LAN (TCP/IP, 10BASE-T/100BASE-TX) RS-232C (Max. 38.4 kbps, Available as printer I/F) EXT I/O (37-pin Handler interface) Analog output (DC 0V to 3.1V) |
Power Supply | 100 to 240V AC, 50Hz/60Hz, 35VA max. |
Dimensions and Mass | 215 mm (W) × 80 mm (H) × 295 mm (D), 2.4 kg |
Included Accessories | Instruction manual ×1, Power cord ×1, Operating Precautions ×1 |